COLLEAGUE
Report
- Report Number
- 6000001-2011-22598
- Event Type
- Malfunction
- Date Received
- September 6, 2011
- Date of Event
- March 2, 2011
- Report Date
- July 18, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BATTERY DEPLETED SET ALARM WAS DISCOVERED AND CONFIRMED BY BAXTER PERSONNEL DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE. A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO PRODUCT DISTRIBUTED WITHIN THE U.S.
THE FACILITY ORIGINALLY REPORTED A COLLEAGUE INFUSION PUMP WITH DAMAGED BATTERIES TO BAXTER CANADA. DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER CANADA PERSONNEL, THE PUMP WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET ALARM, WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.63.92.