LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES
Report
- Report Number
- 3015876-2011-00644
- Event Type
- Death
- Date Received
- August 31, 2011
- Date of Event
- August 2, 2011
- Report Date
- August 2, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K973486
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- UNKNOWN
Narratives
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO OBSERVED A CRITICAL FAULT CODE LOGGED IN THE SERVICE MEMORY, THEN REPLACED THE THERAPY PCB ASSEMBLY. PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING AND THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. THE THERAPY PCB ASSEMBLY WAS DISPOSED OF AND WILL NOT BE RETURNED TO PHYSIO-CONTROL FOR FURTHER EVALUATION. THE ROOT CAUSE OF THE REPORTED FAILURE COULD NOT BE DETERMINED. A CLINICAL EVALUATION OF THE REPORTED EVENT WAS PERFORMED AND DETERMINED THAT THE DEVICE USE MAY HAVE CONTRIBUTED TO THE DEATH OF THE PATIENT BECAUSE DEFIBRILLATION THERAPY WAS DELAYED GREATER THAN 30 SECONDS.
IT WAS REPORTED THAT DURING A PATIENT INCIDENT, THE DEVICE WAS CHARGED UP AND DELIVERED ONE SUCCESSFUL DEFIBRILLATION SHOCK; HOWEVER AFTER THE INITIAL SHOCK, THE END USER ATTEMPTED TO CHARGE THE DEFIBRILLATION ENERGY ONCE AGAIN, BUT THE SERVICE LIGHT ILLUMINATED AND THE DEVICE WOULD NO LONGER PROPERLY CHARGE DEFIBRILLATION ENERGY. A BACK-UP DEVICE WAS BROUGHT IN APPROXIMATELY FOUR (4) MINUTES LATER AND DELIVERED AN ADDITIONAL 8 DEFIBRILLATION SHOCKS. AT THAT POINT, THE HOSPITAL STAFF DECIDED TO CEASE FURTHER DEFIBRILLATION SHOCKS AND CPR. THE PATIENT DID NOT SURVIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES | MKJ | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Death |