FDA Adverse Event Malfunction Summary report: N

12" (30 CM) EXT SET W/1.2 MICRON FILTER, CLAMP, ROTATING LUER

MDR report key: 22340974 · Received June 26, 2025

Report

Report Number
9617594-2025-01354
Event Type
Malfunction
Date Received
June 26, 2025
Date of Event
June 1, 2025
Report Date
August 20, 2025
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
FMG
UDI-DI
00840619045524
PMA / PMN Number
K964435
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR EVALUATION; HOWEVER, TESTING HAS NOT YET BEEN COMPLETED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION D9. RECEIVED TWO (2) USED. LIST # B1115, 12" (30 CM) EXT SET W/1.2 MICRON FILTER, CLAMP, ROTATING LUER FOR INSPECTION. THE TUBING ON THE FILTER OUTLET WAS DISCONNECTED ON ONE OF THE SAMPLES. THE BOND WAS EXAMINED UNDER UV LIGHT AND THERE WAS NO EVIDENCE OF SOLVENT PRESENT. THE TUBING AND FILTER WERE WITHIN DIMENSIONAL SPECIFICATIONS. THE COMPLAINT OF FILTER DISCONNECTING FROM TUBING IS CONFIRMED ON ONE (1) SAMPLE. THE PROBABLE CAUSE OF THE TUBING SEPARATION IS INSUFFICIENT SOLVENT APPLIED DURING MANUAL BOND ASSEMBLY IN MANUFACTURING. THE COMPLAINT OF FILTER DISCONNECTING CANNOT BE REPLICATED OR CONFIRMED ON ONE (1) SAMPLE. A DEVICE HISTORY LOT REVIEW COULD NOT BE CONDUCTED BECAUSE NO LOT NUMBER(S) WAS/WERE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT, ON AN UNSPECIFIED DATE, A 12" (30 CM) EXT SET W/1.2 MICRON FILTER, CLAMP, ROTATING LUER WAS FOUND TO HAVE EXPERIENCED A DISCONNECT AT THE POINT OF THE FILTER. THERE WAS NO PATIENT HARM REPORTED. THE INITIAL REPORTER INDICATED THAT THE EVENT OCCURRED TWICE; HOWEVER, NO ADDITIONAL INFORMATION WAS GIVEN PERTAINING TO THE ADDITIONAL INSTANCES. A SEARCH OF THE COMPLAINT DATABASE SHOWED NO PRIOR REPORT OF THESE EVENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1538075 12" (30 CM) EXT SET W/1.2 MICRON FILTER, CLAMP, ROTATING LUER STOPCOCK, I.V. SET FMG ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. NOT PROVIDED 00840619045524

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown