FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2234037 · Received August 29, 2011

Report

Report Number
3004209178-2011-82743
Event Type
Injury
Date Received
August 29, 2011
Date of Event
August 14, 2011
Report Date
August 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETES KETOACIDOSIS. THE BLOOD GLUCOSE READING WAS 856MG/DL. IT WAS STATED THAT THE CUSTOMER WAS EXPERIENCING UNEXPLAINED HIGH GLUCOSE FOR THE PAST TWO DAYS. TROUBLESHOOTING WAS DECLINED, AND THE CUSTOMER STATED THAT THE DOCTOR BELIEVES THE INSULIN PUMP WAS NOT FUNCTIONING PROPERLY. THE CUSTOMER'S MOST RECENT GLUCOSE READING WAS 358MG/DL, AND HE HAS TREATED WITH MANUAL INJECTIONS. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 17 YR