METASUL LARGE DIAMETER HEAD 42/H
Report
- Report Number
- 9613350-2011-00565
- Event Type
- Injury
- Date Received
- August 30, 2011
- Date of Event
- July 27, 2009
- Report Date
- July 27, 2009
- Manufacturer
- ZIMMER, INC.
- Product Code
- KWA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE MFR DID NOT RECEIVED EXPLANTED DEVICES FOR REVIEW. WHERE LOT NUMBERS WERE RECEIVED FOR THE EXPLANTED DEVICES, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. BASED ON AN EXTENSIVE INVESTIGATION OF EVENTS REPORTED FROM SEVERAL USER FACILITIES OUTSIDE THE U.S., ZIMMER IDENTIFIED THAT THE MOST PROBABLE CAUSE FOR THE OUTCOME OBSERVED WAS A LOOSE OR UNSTABLE CUP THAT RESULTED FROM USE OF SURGICAL TECHNIQUES NOT CONSISTENT WITH THE MFR'S RECOMMENDATIONS. AS A CORRECTIVE ACTION, A RETRAINING PROGRAM FOR USERS OUTSIDE THE U.S. WAS INITIATED IN (B)(6) 2009 AND REPORTED TO THE NATIONAL COMPETENT AUTHORITIES AS REQUIRED. THE DUROM CUP REPORTED IN THIS CASE IS NOT MARKETED IN THE U.S. A SIMILAR CUP, COMPATIBLE WITH THE METASUL LDH FEMORAL HEAD, IS CLEARED IN THE U.S. AND A CORRECTIVE ACTION FOR THIS PRODUCT WAS REPORTED TO THE FDA IN (B)(4) 2008 AS CORRECTION (B)(4). SHOULD ADD'L INFO BECOME AVAILABLE AND / OR THE DEVICE (S) BE RETURNED FOR EVAL AND AN INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE FILED. THE NEED FOR FURTHER CORRECTIVE MEASURES IS NOT INDICATED AT THIS TIME AND ZIMMER GMBH CONSIDERS THIS CASE AS CLOSED. (B)(4).
IT WAS REPORTED THAT THE PT HAD A REVISION SURGERY DUE TO OSTEOLYSIS. POSSIBLE WEAR BETWEEN STEM AND ADAPTER WAS QUESTIONED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | METASUL LARGE DIAMETER HEAD 42/H | DUROM ACETABULAR COMPONENT AND METASUL | KWA | ZIMMER, INC. | 2296581 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Hospitalization| R |