FDA Adverse Event
Malfunction
Summary report: N
IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM
MDR report key: 2233861
·
Received September 4, 2011
Report
- Report Number
- 2050012-2011-05146
- Event Type
- Malfunction
- Date Received
- September 4, 2011
- Date of Event
- August 5, 2011
- Report Date
- August 5, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K962294
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED A "CLOGGED" FILTER IN THE PROBE RINSE WELL VACUUM LINE. THE FSE VERIFIED REPAIRS BY PRIMING SEVERAL TIMES AND MANUALLY ACTIVATING PROBE RINSE. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT WASH SOLUTION WAS OVERFLOWING FROM THE SAMPLE PROBE WASH STATION IN THE IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO ERRORS WERE GENERATED BY THE INSTRUMENT DURING THE LEAKING. NO OPERATOR EXPOSURE WAS NOTED.