FDA Adverse Event Malfunction Summary report: N

IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM

MDR report key: 2233861 · Received September 4, 2011

Report

Report Number
2050012-2011-05146
Event Type
Malfunction
Date Received
September 4, 2011
Date of Event
August 5, 2011
Report Date
August 5, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K962294
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED A "CLOGGED" FILTER IN THE PROBE RINSE WELL VACUUM LINE. THE FSE VERIFIED REPAIRS BY PRIMING SEVERAL TIMES AND MANUALLY ACTIVATING PROBE RINSE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT WASH SOLUTION WAS OVERFLOWING FROM THE SAMPLE PROBE WASH STATION IN THE IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO ERRORS WERE GENERATED BY THE INSTRUMENT DURING THE LEAKING. NO OPERATOR EXPOSURE WAS NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. IMMAGE 800 N/A

Patients

Seq Age Sex Outcome Treatment
1