FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2233827 · Received September 4, 2011

Report

Report Number
2050012-2011-05080
Event Type
Malfunction
Date Received
September 4, 2011
Date of Event
August 4, 2011
Report Date
August 4, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2011, A BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT: FSE EMPTIED THE AIR WATER SEPARATOR AND CLEANED THE WASTE CANISTER. SETTING WAS ADJUSTED ON A WASTE FLOAT SWITCH TO ALLOW FOR PROPER DRAINAGE. REPAIR VERIFIED PER ESTABLISHED PROCEDURES. ISSUE WAS RESOLVED BY A ROUTINE SERVICE REPAIR. BEC INTERNAL NOTIFICATION NUMBER IS (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) CUSTOMER TECHNICAL SUPPORT (CTS) TO REPORT A HYDRO ERROR AND FOUND THE HYDRO DRAWER FULL OF LIQUID. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT DURING THE EVENT. PER CUSTOMER, ABOUT 1 LITER LEAKED ONTO FLOOR. CTS DISCUSSED HOW TO EMPTY THE WASTE CONTAINERS THAT WERE FULL AND THEN HOME THE INSTRUMENT. CTS GENERATED A SERVICE REQUEST. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1