UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2011-05080
- Event Type
- Malfunction
- Date Received
- September 4, 2011
- Date of Event
- August 4, 2011
- Report Date
- August 4, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
ON (B)(4) 2011, A BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT: FSE EMPTIED THE AIR WATER SEPARATOR AND CLEANED THE WASTE CANISTER. SETTING WAS ADJUSTED ON A WASTE FLOAT SWITCH TO ALLOW FOR PROPER DRAINAGE. REPAIR VERIFIED PER ESTABLISHED PROCEDURES. ISSUE WAS RESOLVED BY A ROUTINE SERVICE REPAIR. BEC INTERNAL NOTIFICATION NUMBER IS (B)(4).
THE CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) CUSTOMER TECHNICAL SUPPORT (CTS) TO REPORT A HYDRO ERROR AND FOUND THE HYDRO DRAWER FULL OF LIQUID. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT DURING THE EVENT. PER CUSTOMER, ABOUT 1 LITER LEAKED ONTO FLOOR. CTS DISCUSSED HOW TO EMPTY THE WASTE CONTAINERS THAT WERE FULL AND THEN HOME THE INSTRUMENT. CTS GENERATED A SERVICE REQUEST. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |