FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2233799 · Received September 4, 2011

Report

Report Number
2050012-2011-05079
Event Type
Malfunction
Date Received
September 4, 2011
Date of Event
August 4, 2011
Report Date
August 4, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT: SERVICE HAD BEEN PERFORMED, PER THE CUSTOMER, FOR 1 REAGENT PROBE LEAKING AND CUVETTE NOT DRY ERRORS. FSE REPLACED THE SYRINGE BARREL AND THE 3 WAY VALVE. FSE REPLACED THE T-VALVE TO ADDRESS THE LEAKING. FSE DETERMINED THE CUVETTE NOT DRY ERRORS WERE DUE TO A PLUGGED WASH TOWER PROBE; THE CUSTOMER REPLACED THIS PART. NO FURTHER REPORTS OF LEAKING AND/OR RELATED ERRORS NOTED IN ORACLE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) CUSTOMER TECHNICAL SUPPORT TO REPORT BOTH REAGENT PROBES LEAKED ON THE UNICEL DXC 600I SYNCHRON CHEMISTRY ANALYZER. CUSTOMER STATED THE FITTINGS ON TOP OF THE REAGENT PROBES ARE TIGHT. NO INJURY OR EXPOSURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE. JJE BECKMAN COULTER INC. DXC 600I N/A

Patients

Seq Age Sex Outcome Treatment