FDA Adverse Event Injury Summary report: N

AUTOSOFT XC

MDR report key: 22337652 · Received June 26, 2025

Report

Report Number
3003442380-2025-11043
Event Type
Injury
Date Received
June 26, 2025
Date of Event
June 2, 2025
Report Date
July 14, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6010709 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3.0 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2.0 FOR THE CODE IDD-PMC03.03 OCCLUSION (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). (SOURCE/CAUSE CANNOT BE IDENTIFIED, ONLY USE WHEN A SPECIFIC MALFUNCTION CANNOT BE DETERMINED). COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010709 WAS MANUFACTURED ACCORDING TO THE WI VERSION 120 IN THE LINE 10, ON 11/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. GLUING TUBING: THE LOT 4M00016 WAS ASSEMBLED ACCORDING TO THE FORM VERSION 2 IN THE GLUING OF TUBING IN THE MACHINE ITL03, ON 09/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 07/JUL/2025 AGAINST MALFUNCTION CODE OCCLUSION (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). (SOURCE/CAUSE CANNOT BE IDENTIFIED, ONLY USE WHEN A SPECIFIC MALFUNCTION CANNOT BE DETERMINED) AND LOT 6010709, AND ANOTHER 1 COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT, AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ANOTHER 1 COMPLAINT RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE. NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. BLOOD GLUCOSE LEVEL WAS DISPLAYED HIGH AT THE TIME OF THE EVENT. PATIENT TOOK ASSISTANCE FROM NURSE. PATIENT WAS ALSO SUFFERED FROM THE VOMITING. PATIENT DRANK PLENTY OF FLUIDS. PATIENT CHANGED SOFT CANNULA INFUSION SET ONLY AND RESUMED INSULIN. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1201196 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1001680 6010709 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female