FDA Adverse Event Malfunction Summary report: N

UNICEL® DXL 600 ACCESS® IMMUNOASSAY SYSTEM

MDR report key: 2233750 · Received September 3, 2011

Report

Report Number
2122870-2011-03357
Event Type
Malfunction
Date Received
September 3, 2011
Date of Event
August 8, 2011
Report Date
August 8, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K023764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS DISPATCHED TO THE SITE ON (B)(4) 2011 TO ADDRESS THE ISSUE. THE FIELD SERVICE ENGINEER (FSE) DISCOVERED A "BRIDGE" IN THE INTERLOCK SWITCH. IT APPEARED TO BE A METAL PAPER CLIP AND WAS NOT A BECKMAN COULTER INC. INTERLOCK BYPASS TOOL. THE "BRIDGE" WAS USED TO OVERRIDE THE INTERLOCK CIRCUIT AND ALLOWED THE MOTORS TO BE ENGAGED WHILE THE COVERS WERE OPENED. THE FSE REMOVED THE "BRIDGE" AND THE INTERLOCK SWITCH RETURNED TO NORMAL OPERATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT ON (B)(6) 2011 IT WAS DISCOVERED WHILE PERFORMING MAINTENANCE ON THEIR UNICEL DXL 600 ACCESS IMMUNOASSAY SYSTEM THAT THE INTERLOCK SWITCH FOR THE UNIT WAS NOT DISENGAGING THE MOTORS. THE MOTORS WERE STILL ENGAGED WHILE THE INSTRUMENT COVER WAS OPEN. NO INJURIES WERE REPORTED AND NO PERSONNEL SOUGHT ANY MEDICAL ATTENTION IN ASSOCIATION WITH THIS EVENT. NO PATIENT RESULTS WERE AFFECTED BY THIS EVENT. NO DEATH, INJURY OR MODIFICATION TO PATIENT TREATMENT WAS ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXL 600 ACCESS® IMMUNOASSAY SYSTEM ANALYZER, CHEMISTRY JJE BECKMAN COULTER, INC. NA NA