FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2233740 · Received September 3, 2011

Report

Report Number
1423500-2011-11671
Event Type
Malfunction
Date Received
September 3, 2011
Date of Event
August 15, 2011
Report Date
August 15, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS NOT AVAILABLE AND THE LOT NUMBER IS UNKNOWN. SINCE THE LOT NUMBER IS UNKNOWN, NO BATCH REVIEW WILL BE PERFORMED. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS CONFIRMED DUE TO THE USE ERROR DESCRIBED BY THE PATIENT. THE HOME PATIENT HAD FAILED TO CONNECT TO THE CYCLER WHEN INITIAL DRAIN BEGAN, ALLOWING AIR IN THE LINE. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE POTENTIAL USE ERROR IN THIS COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT A SYSTEM ERROR (SE) 2240 WHICH OCCURRED ON HOME CHOICE (HC) ON DURING USE DURING INITIAL DRAIN. THE HOME PATIENT (HP) STATED THAT SHE DID NOT CONNECT THE PATIENT EXTENSION LINE BEFORE STARTING PRIME. THE HP CYCLED POWER TO CLEAR THE SE 2240 FOLLOWED BY SE 2367 AND ENDED THERAPY. THE BAXTER TECHNICAL REPRESENTATIVE (TSR) REVIEWED PROPER SET UP PROCEDURES. THE TSR HAD THE HP DISCONNECT USING ASEPTIC TECHNIQUE AND REMOVED THE CASSETTE. THE HP WILL NOTIFY THE PERITONEAL DIALYSIS REGISTERED NURSE AND START OVER WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 63 YR