FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2233738 · Received September 3, 2011

Report

Report Number
2531779-2011-06467
Event Type
Malfunction
Date Received
September 3, 2011
Date of Event
August 6, 2011
Report Date
August 7, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(6). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT ON (B)(6) 2011 THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS BECAME INTERMITTENTLY UNRESPONSIVE. SHE STATED THAT THE PATIENT WEARS THE PUMP EXTERIOR TO HIS CLOTHING, AND DOES NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 86 YR