FDA Adverse Event Malfunction Summary report: N

OLYMPUS AU5400 CLINICAL CHEMISTRY ANALYZER

MDR report key: 2233734 · Received September 3, 2011

Report

Report Number
2050012-2011-04753
Event Type
Malfunction
Date Received
September 3, 2011
Date of Event
August 3, 2011
Report Date
August 4, 2011
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
JGS
PMA / PMN Number
K011720
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS DISPATCHED TO THE SITE ON (B)(4) 2011 FOR THIS EVENT. THE FIELD SERVICE ENGINEER (FSE) FOUND A CLOT IN THE INLET BLOCK OF FLOW CELL. THE FSE CLEARED THE CLOT. THE FSE PERFORMED AN ION-SELECTIVE ELECTRODE CHEMISTRY CALIBRATION, QUALITY CONTROL (QC) AND PRECISION ASSESSMENT ON ALL 3 LEVELS OF QC WITH ACCEPTABLE RESULTS. THE INSTRUMENT WAS RETURNED TO SERVICE AFTER THE NECESSARY REPAIRS. THE ROOT CAUSE OF THIS EVENT WAS A CLOGGED INLET BLOCK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT ON (B)(6) 2011 AN ERRONEOUS LOW SODIUM RESULT WAS GENERATED FROM AN OLYMPUS AU5400 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULT WAS REPORTED OUTSIDE THE LABORATORY BUT WAS QUESTIONED BY THE PHYSICIAN. THERE WAS NO DEATH, SERIOUS INJURY OR MODIFICATION TO PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. UPON TEST REPEAT, THE SODIUM REPEAT RESULT WAS HIGHER. NO PATIENT SPECIFIC INFORMATION, SAMPLE HANDLING/COLLECTION INFORMATION OR SYSTEM INFORMATION WAS PROVIDED BY THE CUSTOMER.