UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2011-01360
- Event Type
- Malfunction
- Date Received
- September 3, 2011
- Date of Event
- August 4, 2011
- Report Date
- August 4, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- OTHER
Narratives
SAMPLE COLLECTION AND STORAGE INFORMATION WERE NOT PROVIDED. THE CONTROLS WERE RUN PRIOR TO PROCESSING THE SPECIFIC SPECIMEN AND AFTER THE EVENT. ALL RESULTS WERE WITHIN THE SPECIFICATIONS. THE FLAGGING SENSITIVITY SETTINGS WERE [3303], WHERE THE BLASTS, VARIANT LYMPH, IMM NE 1, IMM NE 2, WERE ALL SET AT HIGHEST LEVEL AND IMM NE1 WAS SET TO OFF. RAW DATA ANALYSIS PROVIDED THE FOLLOWING: THE SAMPLE SHOWS A RELATIVELY SMALL AMOUNT OF NEUTROPHILS. THE NEUTROPHIL POPULATION IS TOUCHING THE LYMPH POPULATION. THE PATTERN IS SLIGHTLY ABNORMAL, BUT IT WAS NOT SIGNIFICANT ENOUGH TO TRIGGER THE BLAST FLAGGING RULES. ON (B)(4) 2011, BEC FIELD SERVICE ENGINEER COLLECTED DATA AND VERIFIED THE INSTRUMENT. THE INSTRUMENT WAS PERFORMING WITHIN THE SPECIFICATIONS. ROOT CAUSE FOR THIS EVENT IS UNKNOWN.
A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM GENERATED ERRONEOUS DIFFERENTIAL RESULTS. THE INSTRUMENT DID NOT GENERATE A DIFFERENTIAL SUSPECT BLAST FLAG FOR A LYMPHOMA PATIENT WITH 28% BLASTS CELLS. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. MANUAL DIFFERENTIAL TEST SHOWED 28% BLASTS CELLS. CORRECTED DIFFERENTIAL REPORT WAS ISSUED BASED ON THE MANUAL TEST. THE PATIENT RESULTS ARE SHOWN. THERE WAS NO DEATH, INJURY, OR EFFECT TO PATIENT TREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | DXH 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |