FDA Adverse Event Malfunction Summary report: N

IMMAGE® IMMUNOCHEMISTRY SYSTEM

MDR report key: 2233723 · Received September 3, 2011

Report

Report Number
2050012-2011-04762
Event Type
Malfunction
Date Received
September 3, 2011
Date of Event
August 3, 2011
Report Date
August 3, 2011
Manufacturer
BECKMAN COUTLER INC.
Product Code
JJE
PMA / PMN Number
K962294
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 FOR THIS EVENT. THE FSE CLEANED THE FRAME AND REMOVED THE ISOLATRAN FROM THE SYSTEM. FRESH CUVETTES WERE PLACED ON THE REACTION WHEEL BECAUSE OF THE EXTENDED DOWN TIME OF THE INSTRUMENT. THE ISOLATRAN AND ASSOCIATED COMPONENTS THAT WERE DAMAGED BY FIRE AND/OR SMOKE WERE REPLACED. UNIT WAS REASSEMBLED AND ALIGNED. QC WAS TESTED WITH NO PROBLEMS NOTED PRIOR TO RETURNING THE INSTRUMENT INTO SERVICE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) THAT THE IMMAGE IMMUNOCHEMISTRY SYSTEM CAUGHT ON FIRE. THE CUSTOMER INDICATED THAT THE FIRE DEPARTMENT WAS DISPATCHED FOR THIS EVENT. THE LABORATORY PERSONNEL WAS EVACUATED AND THE SPRINKLER SYSTEM DID NOT ENGAGE. THERE WERE NO REPORTS OF INJURY AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE® IMMUNOCHEMISTRY SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COUTLER INC. IMMAGE NA

Patients

Seq Age Sex Outcome Treatment
1