FLOGARD
Report
- Report Number
- 6000001-2011-22509
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Date of Event
- August 10, 2011
- Report Date
- August 10, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE SERVICED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE DAMAGED MAIN BATTERIES. TO CORRECT THE CONDITION, THE MAIN BATTERIES WERE REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
(B)(4). THIS DEVICE WAS NOT RETURNED TO BAXTER FOR EVALUATION, HOWEVER, A BAXTER FIELD SERVICE TECHNICIAN REPAIRED THIS DEVICE AT THE CUSTOMER SITE. DEVICE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP REQUIRED REPLACEMENT OF A NP2-N2 BATTERY. "THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT." NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |