INTERMATE
Report
- Report Number
- 6000001-2011-22548
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Date of Event
- August 12, 2011
- Report Date
- August 12, 2011
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- FRN
- PMA / PMN Number
- K910425
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A RUPTURED RESERVOIR WAS CONFIRMED. VISUAL EXAMINATION OF THE UNIT DETERMINED THAT THE BLADDER RUPTURED IN A NON-FOOTED POSITION. APPROXIMATELY 30 ML OF SOLUTION WAS ALSO NOTED IN THE HOUSING DUE TO THE RUPTURE. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION OF THE BLADDER RUPTURE ISSUE IS IN PROGRESS THROUGH (B)(4).
BAXTER SERVICE PERSONNEL REPORTED AN INTERMATE IN WHICH THE RESERVOIR WAS RUPTURED. THIS CONDITION OCCURRED DURING EVALUATION. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTERMATE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - IRVINE | 10B049 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |