FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2233488 · Received September 2, 2011

Report

Report Number
2531779-2011-06459
Event Type
Malfunction
Date Received
September 2, 2011
Report Date
August 4, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX HISTORY SHOWED THAT REBOOTING HAD OCCURRED. THERE WERE NO ALARMS RELATED TO THE COMPLAINT IN THE PUMP ALARM HISTORY. NO DAMAGE WAS FOUND TO BE BATTERY COMPARTMENT OR CAP. THE BATTERY CAP ATTACHED SECURELY TO THE PUMP AND THE PUMP POWERED ON NORMALLY. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ALARMS OR POWER ISSUES. THE BATTERY CAP WAS TESTED AND NO CONNECTION DEFECTS WERE FOUND. THE PUMP WAS OPENED AND NO DAMAGE WAS FOUND TO THE POWER CIRCUIT.

Additional Manufacturer Narrative · 1

AFTER THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT THE ANIMAS PUMP IS CONTINUALLY REBOOTING WHEN SHE ATTEMPTS TO PRIME. THERE WAS NO PRODUCT MISUSE. THE BATTERY CAP AND COMPARTMENT WAS NOT DAMAGED. THE BATTERY COMPARTMENT WAS CORROSION FREE. THE BATTERY WAS REPLACED BUT THE ISSUE WAS NOT RESOLVED WITH TRAINING. THE PRODUCT WAS SENT BACK FOR INVESTIGATION. THERE WAS NO REPORT OF ANY PATIENT IMPACT ASSOCIATED WITH THE COMPLAINT. THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE POWER ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 18 YR