FDA Adverse Event Malfunction Summary report: N

SEE H-10

MDR report key: 2233423 · Received September 2, 2011

Report

Report Number
2023826-2011-00734
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 9, 2011
Report Date
August 10, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P880091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

METHOD: DEVICE HISTORY RECORD REVIEW. RESULTS: A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED AND NOTHING WAS FOUND IN THE MANUFACTURING, PACKAGING, STERILIZATION AND SHIPPING PROCESSES OF THIS LENS THAT WAS THE ROOT CAUSE OF THE COMPLAINT. VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS TORN INTO TWO PIECES. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH, DEVICE HISTORY RECORD REVIEW AND THE EVALUATION OF THE RETURNED PRODUCT, THE MOST LIKELY ROOT CAUSE OF THE EVENT WOULD BE DUE TO MISHANDLING OF THE LENS AT THE FACILITY. (B)(4).

Additional Manufacturer Narrative · 1

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIC) WITH TORIC OPTIC. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS - (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY AND WORK ORDER SEARCH, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. LENS NOT RETURNED.

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON OPENED THE TRAY OF AN AA4203TL TORIC SILICONE SINGLE PIECE LENS AND NOTED THE LENS WAS TORN. THE LENS WAS NOT LOADED OR USED AND THERE WAS NO PATIENT CONTACT. ANOTHER SAME MODEL LENS WAS IMPLANTED. THE REPORTER STATED THE SURGEON INDICATED THE LENS WAS RECEIVED DEFECTIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEE H-10 INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY AA4203TL NA

Patients

Seq Age Sex Outcome Treatment
1 85 YR CARTRIDGE MODEL AND LOT NUMBER UNK| INJECTOR MODEL AND LOT NUMBER UNK