FDA Adverse Event Malfunction Summary report: N

ACCESS

MDR report key: 2233411 · Received September 2, 2011

Report

Report Number
6000001-2011-22387
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
May 13, 2011
Report Date
May 13, 2011
Manufacturer
BAXTER HEALTHCARE - MALTA
Product Code
FPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ONE ACTUAL SAMPLE WAS RETURNED TO THE PLANT FOR EVALUATION AND WAS CONFIRMED THROUGH VISUAL INSPECTION; A TEAR IN THE POUCH PAPER OF THE UNIT WAS OBSERVED. A BATCH REVIEW WAS PERFORMED ON THE REPORTED LOT WITH NO DEVIATIONS FOUND. ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Description of Event or Problem · 1

DURING INSPECTION IN A COLOMBIA WAREHOUSE BEFORE DISTRIBUTION TO A CUSTOMER, 1 UNIT WAS OBSERVED TO HAVE A DAMAGED OVERPOUCH. SAMPLE IS AVAILABLE. PATIENT INJURY AND MEDICAL INTERVENTION WERE NOT REPORTED FOR THIS EVENT. NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE - MALTA N/A 11C31V897

Patients

Seq Age Sex Outcome Treatment
1