FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 2233399 · Received September 2, 2011

Report

Report Number
1423500-2011-11658
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 1, 2011
Report Date
August 16, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: ONE USED SET WAS RECEIVED WITH CAP ON DARK BLUE CONNECTOR AND NO CAP ON PATIENT CONNECTOR IN REFERENCE TO DAMAGED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. THE SET WAS FUNCTIONALLY HAND TIGHTENED TO A LAB TITANIUM ADAPTER WITHOUT DIFFICULTY AND PRESSURE TESTED UNDERWATER WITH LEAK NOTED. SET WAS VISUALLY INSPECTED AND NOTED THAT BOTH OF THE OCCLUDER FEET WAS BROKEN. THIS COMPLAINT WAS CONFIRMED IN THE LAB FOR EXTENSIVE TWIST CLAMP CRACKING INCLUDING BROKEN OCCLUDER FEET. A ROOT CAUSE IS UNCERTAIN. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THIS IS REPORT 1 OF 3 ASSOCIATED WITH THIS ISSUE. THE SAMPLE HAS BEEN REPORTED TO BE AVAILABLE FOR EVALUATION AND HAS BEEN REQUESTED. A FOLLOW-UP REPORT WILL BE FILED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER TO REPORT THAT THREE TRANSFER SETS WERE NOTED TO HAVE SOME CRACKS ON THE LIGHT BLUE MAIN BODY. THESE SETS WERE USED FOR ONE MONTH. THERE WERE PATIENTS INVOLVEMENT. BUT NO PATIENT INJURY AND NO MEDICAL INTERVENTION WAS REPORTED ALONG WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H11B21041

Patients

Seq Age Sex Outcome Treatment
1