FDA Adverse Event Malfunction Summary report: N

AUTOSOFT XC

MDR report key: 22333912 · Received June 26, 2025

Report

Report Number
3003442380-2025-11014
Event Type
Malfunction
Date Received
June 26, 2025
Date of Event
May 29, 2025
Report Date
September 2, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF 3709030 DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF 3709030. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD 2304856. THE BATCH 6010109, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. · DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010109 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 118 AND MANUFACTURED IN THE LINE 10, ON 07/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 4K05987 WAS MANUFACTURED ACCORDING TO THE FORM VERSION 2 AND MANUFACTURED IN THE MACHINE ITL03, ON 05/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED INFUSION SET DETACHMENT EVENT ON (B)(6) 2024. THE SITE OF DETACHMENT WAS FROM HUB. THE INFUSION SET WAS IN USE FOR 1MINUTE. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1792778 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1001680 6010109 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 13 YR Male