FDA Adverse Event Malfunction Summary report: N

APOLLO

MDR report key: 2233358 · Received August 31, 2011

Report

Report Number
2233358
Event Type
Malfunction
Date Received
August 31, 2011
Date of Event
August 30, 2011
Report Date
August 31, 2011
Manufacturer
DRAEGER GMBH
Product Code
BSZ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US

Narratives

Description of Event or Problem · 1

THE DRAEGER APOLLO ANESTHESIA MACHINE HAS THE PT. MONITOR MOUNTED TO AN ARM THAT IS ATTACHED TO THE MAIN BODY OF THE ANESTHESIA-GAS MACHINE BY SCREWS ATTACHED TO A TWO AND SEVEN-EIGHTH INCH SQUARE ALUMINUM PIECE IN A TRACK HELD IN BY M6 SCREWS. BOTH ALUMINUM PIECES FAILED BECAUSE OF THREAD FAILURE. THE ALUMINUM THREADS WERE STILL ATTACHED TO THE SCREWS WHEN THEY WERE RECOVERED FROM THE FLOOR. THIS IS THE SECOND OCCURENCE OF THIS PROBLEM.====================== MANUFACTURER RESPONSE FOR ANESTHESIA MACHINE/MONITOR, APOLLO (PER SITE REPORTER)======================THEY WERE CONTACTED FOR THE FIRST EVENT. THEY COULD NOT LOCATE A PART NUMBER OR THE CORRECT PART FOR THE FAILED ALUMINUM PIECE. I WILL CALL THEIR QUALITY CONTROL DEPARTMENT ON THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 APOLLO GAS-MACHINE, ANESTHESIA BSZ DRAEGER GMBH * *

Patients

Seq Age Sex Outcome Treatment
1 *