FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2233243 · Received September 2, 2011

Report

Report Number
1644487-2011-02065
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
March 18, 2010
Report Date
August 12, 2011
Manufacturer
CYBERONICS
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING MANUFACTURER REVIEW OF A PATIENT'S VNS PROGRAMMING HISTORY, IT WAS NOTED A FAULTED SYSTEMS DIAGNOSTICS TEST OCCURRED ON (B)(6) 2010 WHICH CAUSED THE PATIENT'S SETTING TO CHANGE. THE SETTING CHANGE WAS NOTED UPON REINTERROGATION OF THE VNS, BUT ONLY THE OUTPUT CURRENT WAS CORRECTED. THE OFF TIME WAS LEFT AT 60 MINUTES. THE SETTINGS WERE NOT CORRECTED UNTIL ON OR BEFORE (B)(6) 2010 PER THE HISTORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS MODEL 250 NI

Patients

Seq Age Sex Outcome Treatment
1 11 YR