FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL

MDR report key: 2233195 · Received July 22, 2011

Report

Report Number
1811755-2011-02653
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 24, 2011
Report Date
June 24, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY THE MFR AND THE REPORTED FAILURE BY THE CUSTOMER WAS DUPLICATED. UPON VISUAL EXAMINATION, IT WAS FOUND THAT THE MOTOR INLET HOSE WAS RIPPED APART FROM THE SWIVEL ASSEMBLY. THE MOTOR INLET HOSE CONNECTS TO THE SWIVEL ASSEMBLY; HIGH PRESSURE GAS RUNS THROUGH THE INLET HOSE AND IF NOT FITTED CORRECTLY CAN CAUSE THE MOTOR HOSE ASSEMBLY TO DISASSEMBLE OR BLOW APART FROM THE SWIVEL ASSEMBLY. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HEAD OF THE DRILL CASING CAME OFF DURING A CASE. A BACKUP WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO MEDICAL INTERVENTION REQUIRED AND NO DELAY IN THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO DRILL HBB STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK