FDA Adverse Event Malfunction Summary report: N

8MM MCS TIP COVER ACCESSORY

MDR report key: 2233190 · Received September 2, 2011

Report

Report Number
2955842-2011-00289
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
July 12, 2011
Report Date
August 10, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE TIP COVER ACCESSORY WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING FOUND VARIOUS HOLES BURNED THROUGH THE SILICONE. THE SILICONE EXHIBITS LOCALIZED MELTING AROUND THE HOLES, INDICATIVE OF ARCING. THE ULTEM SLEEVE WAS FOUND TO BE MELTED AS ONE HOLE IS LOCATED AT THE INTERFACE BETWEEN THE SILICONE AND ULTEM SLEEVE. THE HOLE SIZES RANGE FROM .140 INCHES TO UNDER .010 INCHES AND ARE LOCATED FROM 0 INCHES TO .300 INCHES ABOVE THE SILICONE TO SLEEVE INTERFACE. THE TIP COVER ACCESSORY WAS ALSO FOUND TO HAVE MECHANICAL TEARS IN THE VICINITY OF THE HOLES, INCLUDING ONE TEAR THAT CONNECTS TWO HOLES.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE, ARCING WAS OBSERVED COMING FROM THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WITH A TIP COVER ACCESSORY INSTALLED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8MM MCS TIP COVER ACCESSORY ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 400180-10 C11136

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS, ACCESSORIES, & ESU