FDA Adverse Event
Malfunction
Summary report: N
UNIVERSAL DRILL
MDR report key: 2233177
·
Received July 22, 2011
Report
- Report Number
- 1811755-2011-02625
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 28, 2011
- Report Date
- July 5, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS REC'D AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS CORROSION AND PROBLEMS WITH THE ROTOR, MOTOR, AND BEARINGS, WHICH WERE EACH REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DRILL BEGAN OVERHEATING WHILE THE EQUIPMENT WAS BEING TESTED. THERE WAS NO PT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNIVERSAL DRILL | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |