FDA Adverse Event
Malfunction
Summary report: N
UNIVERSAL DRILL
MDR report key: 2233172
·
Received July 22, 2011
Report
- Report Number
- 1811755-2011-02630
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 23, 2011
- Report Date
- June 23, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED AND THE REPORTED FAILURE BY THE CUSTOMER WAS CONFIRMED. UPON VISUAL INSPECTION THE BEARINGS ON THE ROTOR ASSEMBLY WERE FOUND TO HAVE FAILED AS A RESULT OF CORROSION DAMAGE. THE BEARINGS FUNCTION TO REDUCE FRICTION DURING ROTOR ROTATION IN THE MOTOR COILS; CORROSION BUILDUP CAN PREVENT THE BEARINGS FROM REDUCING THE FRICTION AND CAN RESULT IN HEAT GENERATION. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DRILL WAS OVERHEATING DURING A PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT. THE PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNIVERSAL DRILL | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |