FDA Adverse Event Malfunction Summary report: N

UNIVERSAL DRILL

MDR report key: 2233172 · Received July 22, 2011

Report

Report Number
1811755-2011-02630
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 23, 2011
Report Date
June 23, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED AND THE REPORTED FAILURE BY THE CUSTOMER WAS CONFIRMED. UPON VISUAL INSPECTION THE BEARINGS ON THE ROTOR ASSEMBLY WERE FOUND TO HAVE FAILED AS A RESULT OF CORROSION DAMAGE. THE BEARINGS FUNCTION TO REDUCE FRICTION DURING ROTOR ROTATION IN THE MOTOR COILS; CORROSION BUILDUP CAN PREVENT THE BEARINGS FROM REDUCING THE FRICTION AND CAN RESULT IN HEAT GENERATION. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DRILL WAS OVERHEATING DURING A PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT. THE PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNIVERSAL DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK