FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2233161 · Received July 22, 2011

Report

Report Number
1811755-2011-02637
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 23, 2011
Report Date
June 23, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

WIDE SPREAD CORROSION DAMAGE WITHIN THE INTERNAL COMPONENTS WAS IDENTIFIED AS THE PROBABLE CAUSE OF THE OVERHEATING DESCRIBED.

Description of Event or Problem · 1

THE CORE IMPACTION DRILL WAS SENT IN FOR EVAL DUE TO OVERHEATING DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A READILY AVAILABLE REPLACEMENT DEVICE WITHOUT ANY CLINICALLY SIGNIFICANT DELAY. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK