FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL

MDR report key: 2233155 · Received July 22, 2011

Report

Report Number
1811755-2011-02622
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 22, 2011
Report Date
June 22, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SERIAL NUMBER ON THE DEVICE WAS NOT READABLE SO THE MANUFACTURER DATE CANNOT BE DETERMINED. THE DEVICE WAS RECEIVED BY THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. A VISUAL EXAMINATION OF THE DRILL CONFIRMED THAT THE PNEUMATIC HOSE HAD A HOLE IN IT. THE HOSE ASSEMBLY WAS REPLACED, AND ADD'L PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE DEVICE WAS RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A HOLE IN THE DRILL'S HOSE BY THE CONNECTOR. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAESTRO DRILL HBB STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK