FDA Adverse Event
Malfunction
Summary report: N
MAESTRO DRILL
MDR report key: 2233155
·
Received July 22, 2011
Report
- Report Number
- 1811755-2011-02622
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 22, 2011
- Report Date
- June 22, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE SERIAL NUMBER ON THE DEVICE WAS NOT READABLE SO THE MANUFACTURER DATE CANNOT BE DETERMINED. THE DEVICE WAS RECEIVED BY THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. A VISUAL EXAMINATION OF THE DRILL CONFIRMED THAT THE PNEUMATIC HOSE HAD A HOLE IN IT. THE HOSE ASSEMBLY WAS REPLACED, AND ADD'L PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE DEVICE WAS RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THERE WAS A HOLE IN THE DRILL'S HOSE BY THE CONNECTOR. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAESTRO DRILL | HBB | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |