SYRINGE 5ML SALINE FILL CHINA SP
Report
- Report Number
- 1911916-2025-00437
- Event Type
- Malfunction
- Date Received
- June 25, 2025
- Date of Event
- May 29, 2025
- Report Date
- July 9, 2025
- Manufacturer
- BECTON DICKINSON
- Product Code
- NGT
- UDI-DI
- 00382903065943
- PMA / PMN Number
- K161552
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
(B)(4) FOLLOW UP. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 306594 AND LOT NUMBER 4142114. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT SIMILAR ISSUES DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.
A NURSE TOOK A 5ML PREFILLED CATHETER FLOSSER TO SEAL THE TUBING, AND WHEN SHE USED IT, SHE FOUND THAT THE CONNECTOR WAS LEAKING, AND SHE FOUND THAT THE SPIRAL CONNECTOR OF THE CATHETER FLOSSER WAS BROKEN, AND SHE IMMEDIATELY REPLACED IT WITH A NEW ONE TO RESEAL THE TUBING, WHICH DID NOT RESULT IN ANY DISCOMFORT OR CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1142661 | SYRINGE 5ML SALINE FILL CHINA SP | SALINE, VASCULAR ACCESS FLUSH | NGT | BECTON DICKINSON | 4142114 | 00382903065943 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |