FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 2232977
·
Received July 22, 2011
Report
- Report Number
- 1811605-2011-00085
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- July 1, 2011
- Report Date
- July 13, 2011
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED THE SWITCH SPARKED. EXAMINATION OF THE UNIT REVEALED A BREAK IN THE CORD AT THE ENTRANCE TO THE STRAIN RELIEF AREA NEAR THE SWITCH. THE BREAK COULD POSSIBLY EXPOSE BARE WIRES WHEN THE CORD IS BENT. THE SWITCH HAD ALSO BEEN DISASSEMBLED BY THE CONSUMER IN AN ATTEMPT TO REPAIR THE UNIT. THE FAILURE MAY BE A RESULT OF MISUSE OR FAILURE TO FOLLOW INSTRUCTIONS ON THE PART OF THE CONSUMER BY SUBJECTING THE CORD TO EXCESSIVE BENDING. ALTHOUGH USERS ARE INSTRUCTED NOT TO BEND THE CORD NEAR THE SWITCH, WE ARE TALKING APPROPRIATE ACTION TO MAKE THE DESIGN MORE ROBUST BY INITIATING A CAPA PROJECT ((B)(4)) TO PREVENT THIS FAILURE FROM RECURRING.
Description of Event or Problem · 1
IT WAS REPORTED THE UNIT STOPPED WORKING.