FDA Adverse Event Malfunction Summary report: N

THERMOPHORE ARTHRITIS PAD

MDR report key: 2232975 · Received July 22, 2011

Report

Report Number
1811605-2011-00083
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 1, 2011
Report Date
June 30, 2011
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THE SWITCH EMITTED SPARKS. UPON EXAMINATION, WE DISCOVERED THAT THE SWITCH WAS NON-FUNCTIONAL WITH CONNECTING WIRES PULLED FROM THE CIRCUIT BOARD, CAUSING A COMPONENT TO SHORT AND EMIT A SPARK. THERE WAS ALSO EVIDENCE THAT THE PAD HAD BEEN FOLDED AND BUNCHED, WHICH IS CONTRARY TO OUR INSTRUCTIONS AND WARNINGS. AT OUR REQUEST, THE MANUFACTURER OF THE SWITCH HAS ENACTED SEVERAL CAPA PROJECTS TO IMPROVE THE DURABILITY OF THE SWITCH. THE CONDITION OF THE PAD SUGGESTS THAT THE PAD MIGHT HAVE BEEN CAUGHT IN SOMETHING WHICH APPLIED AN EXCESSIVE FORCE TO THE CORD.

Description of Event or Problem · 1

IT WAS REPORTED THE SWITCH EMITTED SPARKS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE ARTHRITIS PAD ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 155