FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE ARTHRITIS PAD
MDR report key: 2232975
·
Received July 22, 2011
Report
- Report Number
- 1811605-2011-00083
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 30, 2011
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED THE SWITCH EMITTED SPARKS. UPON EXAMINATION, WE DISCOVERED THAT THE SWITCH WAS NON-FUNCTIONAL WITH CONNECTING WIRES PULLED FROM THE CIRCUIT BOARD, CAUSING A COMPONENT TO SHORT AND EMIT A SPARK. THERE WAS ALSO EVIDENCE THAT THE PAD HAD BEEN FOLDED AND BUNCHED, WHICH IS CONTRARY TO OUR INSTRUCTIONS AND WARNINGS. AT OUR REQUEST, THE MANUFACTURER OF THE SWITCH HAS ENACTED SEVERAL CAPA PROJECTS TO IMPROVE THE DURABILITY OF THE SWITCH. THE CONDITION OF THE PAD SUGGESTS THAT THE PAD MIGHT HAVE BEEN CAUGHT IN SOMETHING WHICH APPLIED AN EXCESSIVE FORCE TO THE CORD.
Description of Event or Problem · 1
IT WAS REPORTED THE SWITCH EMITTED SPARKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE ARTHRITIS PAD | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 155 |