EVIS EXERA II XENON LIGHT SOURCE
Report
- Report Number
- 8010047-2011-00169
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 29, 2011
- Report Date
- June 29, 2011
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- GCT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NURSE
Narratives
OLYMPUS F/U WITH THE USER FACILITY TO OBTAIN ADD'L INFO REGARDING THIS REPORT. THE USER FACILITY REPORTED THAT THE REPORTED PHENOMENON OCCURRED TOWARD THE END OF THE PROCEDURE WHEN THE ENDOSCOPE WAS BEING WITHDRAWN FROM THE CECUM. THE PROCEDURE WAS REPORTEDLY ABORTED. THE DEVICE REFERENCED IN THIS REPORT WAS RETURNED TO OLYMPUS FOR EVAL, HOWEVER, THE MAIN XENON LAMP WAS NOT RETURNED WITH THE UNIT. THE UNIT WAS OPERATED FOR SEVERAL HOURS WITH THE MAIN XENON LAMP POWERED-ON, AND THE DEVICE OPERATED APPROPRIATELY. THE LIGHT SOURCE WAS ABLE TO IGNITE THE XENON LAMP WITHOUT ANY ISSUES, AND THE SPARE LAMP WAS CONFIRMED TO OPERATE APPROPRIATELY. FURTHER INSPECTION NOTED THAT THE CONVERTER FUSE WAS DISCOLORED AND THERE WAS EVIDENCE OF A HIGH TEMPERATURE EVENT, HOWEVER, IT COULD NOT BE DETERMINED IF THIS FINDING WAS ASSOCIATED WITH THE REPORTED EVENT. THE CONVERTER FUSE WAS DISCOLORED AND BORE EVIDENCE OF HIGH HEAT. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
OLYMPUS WAS INFORMED THAT TOWARD THE END OF AN UNSPECIFIED COLONOSCOPY PROCEDURE, THE SUBJECT DEVICE SHUT DOWN COMPLETELY, AND THE USERS EXPERIENCED A COMPLETE LOSS OF IMAGE. THERE WAS NO PT HARM REPORTED, HOWEVER, THE PROCEDURE WAS SAID TO HAVE BEEN ABORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EVIS EXERA II XENON LIGHT SOURCE | LIGHT SOURCE | GCT | OLYMPUS MEDICAL SYSTEMS CORPORATION | CLV-180 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNIDENTIFIED MODEL OF OLYMPUS COLONOSCOPE |