FDA Adverse Event Malfunction Summary report: N

AXIOM ARISTOS FX

MDR report key: 2232970 · Received July 22, 2011

Report

Report Number
2240869-2011-00027
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 24, 2011
Report Date
July 20, 2011
Manufacturer
SIEMENS AG
Product Code
KPR
PMA / PMN Number
K013826
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A DETAILED INVESTIGATION OF THE SYSTEM IS NECESSARY TO DETERMINE THE ROOT CAUSE OF THE DESCRIBED ISSUE. FURTHER INFO WAS REQUESTED. SIEMENS LOCAL SERVICE ENGINEER, (B)(4), INSTRUCTED THE CUSTOMER ON HOW TO USE THE SYSTEM. IN AGREEMENT WITH THE CUSTOMER WARNING LABEL WITH SAFETY INSTRUCTIONS WILL BE POSTED IN THE EXAM ROOM. THE SYSTEM IS EQUIPPED WITH BOTH AN EMERGENCY SHUT-DOWN SWITCH AND TABLE STOP BUTTONS. THESE BUTTONS WILL HALT ALL SYSTEM MOVEMENTS UPON REQUEST AND ARE AVAILABLE TO THE OPERATOR IN CASES OF UNINTENDED DEVICE MOVEMENT. THE REPORTED EVENT APPEARS TO BE A SINGLE OCCURRENCE AND NO SIMILAR ISSUES HAVE REPORTED FROM THE FIELD. CUSTOMER ADDRESS: (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO AN EXAM, THE CUSTOMER SELECTED A PT AND AN ANATOMIC PROGRAM "KNEE FACE ON TABLE". THE PT WAS NOT YET POSITIONED ON THE TABLE, BUT WAS ON A PT BED NEAR THE TABLE TOP. WHEN THE USER GRABBED THE HAND CONTROLLER, WITHOUT PRESSING ANY BUTTONS, THE X-RAY TUBE AND DETECTOR SUDDENLY MOVED AT A VERY FAST SPEED AND DIDN'T STOP, EVEN AFTER THE USER RELEASED THE HAND CONTROLLER. THE X-RAY TUBE STOPPED VERY CLOSE TO THE PT'S HEAD. THE USER DID NOT PUSH THE EMERGENCY STOP BUTTON. THE CUSTOMER ADMITTED THAT THE OPERATOR WAS NOT LOOKING AT THE PT DURING THE SYSTEM MOVEMENTS. THERE IS NO INJURY INVOLVED IN THIS EVENT. THE REPORTED EVENT OCCURRED IN (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AXIOM ARISTOS FX SYSTEM, X-RAY, STATIONARY KPR SIEMENS AG 10093864

Patients

Seq Age Sex Outcome Treatment
1 Other