LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2011-00269
- Event Type
- Malfunction
- Date Received
- July 20, 2011
- Date of Event
- May 25, 2011
- Report Date
- July 20, 2011
- Manufacturer
- ZOLL LIFECOR CORP
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER UP) HAS BEEN CONFIRMED. THE CAUSE OF THE MONITOR NOT POWERING UP HAS BEEN ISOLATED TO AN INTERMITTENT CONNECTION AT PIN 25 OF THE FLASH MEMORY. THE INTERMITTENT CONNECTION RESULTED IN THE FLASH MEMORY BECOMING CORRUPT. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FORM THE CORRUPT MEMORY. THE PT RECEIVED A REPLACEMENT MONITOR.
THE WIFE OF A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PT'S MONITOR WAS "EMITTING A LOUD SCREECHING NOISE WHEN BATTERY WAS PLACED IN IT. [NOW] NEITHER BATTERY WILL POWER IT UP." THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORP | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |