LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2011-00282
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Date of Event
- June 19, 2011
- Report Date
- July 22, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY/CHARGER FAULTS) HAS BEEN CONFIRMED. UPON EVALUATION, THE BATTERY OUTPUT VOLTAGE WAS MEASURED AT 2.84V. THE CAUSE FOR THE LOW OUTPUT VOLTAGE CANNOT BE POSITIVELY DETERMINED. A ROOT CAUSE ANALYSIS IS CURRENTLY UNDERWAY. A SUPPLEMENTAL REPORT WILL BE SENT UPON COMPLETION OF ROOT CAUSE ANALYSIS. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PATIENT RECEIVED A REPLACEMENT BATTERY PACK. NO PROBLEMS WERE DISCOVERED WITH THE PATIENT'S SECOND BATTERY PACK.
THE WIFE OF A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT'S BATTERIES WILL NOT CHARGE. THE PATIENT WAS ISSUED 2 REPLACEMENT BATTERY PACKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR |