FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2232918 · Received July 21, 2011

Report

Report Number
3008642652-2011-00271
Event Type
Malfunction
Date Received
July 21, 2011
Date of Event
June 6, 2011
Report Date
July 20, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE TRUNK CABLE CONNECTOR WAS DAMAGED AND THE CABLE RUNNING FROM THE DISTRIBUTION NODE TO ECG B WAS PULLED FROM THE STRAIN RELIEF. THE ROOT CAUSE OF THE DAMAGED BELT CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

DURING THE INVESTIGATION OF ELECTRODE BELT SN (B)(4) FOR AN UNRELATED COMPLAINT, A REPORTABLE PROBLEM WAS DETECTED DURING SERVICING. THE DISTRIBUTION NODE TO ECG B CABLE WAS PULLED FROM THE STRAIN RELIEF, AND THE TRUNK CABLE WAS DAMAGED. THE PATIENT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 59 YR