LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2011-00271
- Event Type
- Malfunction
- Date Received
- July 21, 2011
- Date of Event
- June 6, 2011
- Report Date
- July 20, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE TRUNK CABLE CONNECTOR WAS DAMAGED AND THE CABLE RUNNING FROM THE DISTRIBUTION NODE TO ECG B WAS PULLED FROM THE STRAIN RELIEF. THE ROOT CAUSE OF THE DAMAGED BELT CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
DURING THE INVESTIGATION OF ELECTRODE BELT SN (B)(4) FOR AN UNRELATED COMPLAINT, A REPORTABLE PROBLEM WAS DETECTED DURING SERVICING. THE DISTRIBUTION NODE TO ECG B CABLE WAS PULLED FROM THE STRAIN RELIEF, AND THE TRUNK CABLE WAS DAMAGED. THE PATIENT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |