FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2232890 · Received September 2, 2011

Report

Report Number
1423500-2011-11625
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
June 13, 2011
Report Date
August 9, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A SYSTEM ERROR (SE) 2240 WAS FOUND DURING A REVIEW OF THE EVENT LOGS OF THE HC DEVICE. NOT ENOUGH DATA IS AVAILABLE WITHIN THE COMPLAINT TO IDENTIFY ROOT CAUSE. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. BAXTER HAS FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). A SAMPLE IS NOT AVAILABLE. A FOLLOW UP REPORT WILL BE SENT WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE EVENT WAS FOUND DURING REVIEW OF THE HOME CHOICE (HC) DEVICE ON (B)(4). THE SYSTEM ERROR OCCURRED ON (B)(6) 2011 @ 10:28 NOT FOUND IN ANY EVENT LOGS. A SE 2240 IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. IT IS UNKNOWN IF THIS ALARM OCCURRED DURING PATIENT THERAPY, HOWEVER NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1