FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8 ML

MDR report key: 2232656 · Received August 24, 2011

Report

Report Number
3004209178-2011-82662
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 8, 2011
Report Date
August 8, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED USED RESERVOIR AND PERFORMED MANUAL PRIME AND HIGH-PRESSURE TESTS. THE RESERVOIR PASSED THE TESTS AND NO LEAKS OR DEFECTS IN THE O-RINGS WERE OBSERVED DURING ANALYSIS. THE RESERVOIR WAS CONNECTED AND LOCKED IN PLACE PROPERLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN FROM BACK OF THE RESERVOIR BARREL WAS LEAKING INTO THE INSULIN PUMP. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESERVOIR 1.8 ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-326A H7762363

Patients

Seq Age Sex Outcome Treatment
1 43 YR