FDA Adverse Event Injury Summary report: N

MINICAP TRANSFER SET

MDR report key: 2232588 · Received September 2, 2011

Report

Report Number
1423500-2011-11620
Event Type
Injury
Date Received
September 2, 2011
Date of Event
July 1, 2011
Report Date
August 15, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE DATE OF ONSET OF THIS PERITONITIS EPISODE IS UNKNOWN THE SAMPLE WAS NOT REQUEST. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS IS REPORT 3 OF 3 INVOLVED IN THIS PERITONITIS INCIDENT.

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WAS CONDUCTED FOR POTENTIALLY ASSOCIATED LOT NUMBERS H10K12043 AND H11F03065 WITH NO EXCEPTIONS OBSERVED RELATED TO THE REPORTED CONDITION. THE COMPLAINT WAS NOT CONFIRMED. NO DEVICE MALFUNCTION OR USER ERROR WAS IDENTIFIED. THE CAUSE OF THE PERITONITIS WAS UNDETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THIS REPORT WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE (GPV) AND IS A SPONTANEOUS REPORT BY A CONSUMER FROM THE USA OF PERITONITIS WITH CULTURE POSITIVE FOR GRAM NEGATIVE ORGANISM IN A PATIENT COINCIDENT WITH DIANEAL PD4 AMBUFLEX THERAPY FOR PERITONEAL DIALYSIS (PD). DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE FOLLOWING WAS REPORTED BY THE CONSUMER. ON (B)(6) 2011, THE PATIENT EXPERIENCED PERITONITIS AND WAS HOSPITALIZED THE SAME DAY. THE CAUSE OF THE PERITONITIS WAS UNKNOWN. TREATMENT WAS NOT REPORTED. THE PATIENT WAS RECOVERING FROM THE PERITONITIS AND REMAINED HOSPITALIZED. ON AN UNREPORTED DATE, THE CATHETER WAS REMOVED AND DIANEAL THERAPY WAS WITHDRAWN. AN OPINION OF CAUSALITY WAS NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 54 YR Hospitalization HOMECHOICE| DIANEAL PD4 AMBUFLEX