IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM
Report
- Report Number
- 0002023141-2025-01708
- Event Type
- Injury
- Date Received
- June 24, 2025
- Date of Event
- May 8, 2025
- Report Date
- August 21, 2025
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019560
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER K011028, K013227.
ZIMVIE RECEIVED ONE (1) TSVH10 (IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM) FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED. FRACTURE IDENTIFIED AT THE COLLAR. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1253960. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1253960 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿DENTAL: FUNCTIONAL: FRACTURE: IMPLANT.¿ THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW AS PER (B)(4) REV 5, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION HAS OCCURRED. THE IMPLANT WAS FRACTURED AT THE COLLAR. THE REPORTED EVENT IS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. D9: DEVICE AVAILABILITY. G3: DATE RECEIVED BY MANUFACTURER. G6: CHECKED "FOLLOW-UP". H2: CHECKED FOLLOW-UP TYPE. H3: CHANGED "NO" TO "YES". H6: ENTERED EVALUATION CODES. H10: ADDED MANUFACTURER NARRATIVE.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 WAS REMOVED AS IT WAS FRACTURED. PATIENT PRESENTED FOR IMPLANT REMOVAL OF IMPLANT. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION, PAIN & SWELLING.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1390204 | IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1253960 | 00889024019560 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Unknown | Required Intervention |