PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2232450
·
Received August 25, 2011
Report
- Report Number
- 2032227-2011-02177
- Event Type
- Injury
- Date Received
- August 25, 2011
- Date of Event
- August 15, 2011
- Report Date
- August 15, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED WITH A HIGH BLOOD GLUCOSE READING OF 547 MG/DL. THE CUSTOMER WAS ALSO GETTING NO DELIVERY ALARMS PRIOR TO THE EVENT. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT THE HIGH PRESSURE TEST COULD NOT BE CONDUCTED AT THE TIME OF THE CALL. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC MINIMED | MMT-723RNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Hospitalization |