FDA Adverse Event Injury Summary report: N

ASR ACETABULAR CUPS 56

MDR report key: 2232381 · Received August 18, 2011

Report

Report Number
1818910-2011-15744
Event Type
Injury
Date Received
August 18, 2011
Date of Event
July 25, 2011
Report Date
October 3, 2016
Manufacturer
DEPUY INTERNATIONAL LTD. 8010379
Product Code
KWA
PMA / PMN Number
K040627
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED.

Additional Manufacturer Narrative · 1

THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN (B)(4) 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADD'L INFO BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. REF. WWCAPA (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THE INVESTIGATION WILL BE REOPENED.

Additional Manufacturer Narrative · 1

NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

PT WAS REVISED TO ADDRESS ACETABULAR LOOSENING. THE CUP HAD ROTATED.

Description of Event or Problem · 1

UPDATE - 10/03/2016 MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS FOR MDR REPORTABILITY, IN ADDITION TO WHAT WAS PREVIOUSLY REPORTED,THE MEDICAL RECORDS REPORT PAIN AND STIFFNESS PART NUMBER FOR HEAD PROVIDED.

Description of Event or Problem · 1

ON 1/9/2017 MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS FOR MDR REPORTABILITY, THE REVISING SURGEON NOTED IN THE SURGICAL INDICATIONS THAT THE PATIENT HAD AN ELEVATED COBALT ION LEVEL OF 17 PPB, ALTHOUGH CHROMIUM THE LEVEL WERE NORMAL. ELEVATED METAL IONS HARM ADDED TO COMPLAINT. ASR SLEEVE AND STEM ADDED TO COMPLAINT. IMPLANT STICKERS ARE AVAILABLE WITHIN THE RECORD BUT ARE FOR INTERNATIONALLY DISTRIBUTED PRODUCTS NEVER SOLD IN THE USA. ACCESS TO MANUFACTURING LOTS, DATES AND POST-MARKET CODES ARE STILL PENDING FOR THOSE PRODUCTS, AND THE COMPLAINT WILL BE UPDATED WHEN/IF THEY BECOME AVAILABLE.

Description of Event or Problem · 1

UPDATE OCT 2, 2017: MEDICAL RECORD RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS FOR THE MDR REPORTABILITY, IN ADDITION TO WHAT WAS PREVIOUSLY REPORTED, REVISION NOTES REPORTED SEVERAL AREAS WERE CYSTIC. PRODUCT CODES AND LOT NUMBERS UPDATED. THIS COMPLAINT WAS UPDATED ON OCT 23, 2017.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASR ACETABULAR CUPS 56 87KWA KWA DEPUY INTERNATIONAL LTD. 8010379 NA 1184624

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention