ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-06427
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Report Date
- August 5, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(4).
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED THAT REBOOTING HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING. VISUAL INSPECTION CONFIRMED NO DAMAGE TO THE BATTERY CAP AND COMPARTMENT; THE BATTERY CAP WAS ABLE TO ATTACH SECURELY TO THE PUMP. THE PUMP POWERS ON NORMALLY WITH NO ALARMS, AND THERE WAS NO DAMAGE FOUND TO THE POWER CIRCUIT. NO DEFECTS WERE FOUND TO THE BATTERY CAP CONNECTIONS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OCCURRING.
A FAMILY MEMBER REPORTED THAT HE NOTICED THE PUMP SCREEN WAS BLANK. HE TOOK THE BATTERY OUT OF THE PUMP AND ATTEMPTED TO REBOOT; THE PUMP BEEPED BUT THE SCREEN REMAINED BLANK. HE TRIED TO INSERT A BATTERY FROM A DIFFERENT PACK AND AGAIN THE PUMP BEEPED BUT THE SCREEN REMAINED BLANK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |