FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2232199 · Received September 2, 2011

Report

Report Number
2531779-2011-06427
Event Type
Malfunction
Date Received
September 2, 2011
Report Date
August 5, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED THAT REBOOTING HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING. VISUAL INSPECTION CONFIRMED NO DAMAGE TO THE BATTERY CAP AND COMPARTMENT; THE BATTERY CAP WAS ABLE TO ATTACH SECURELY TO THE PUMP. THE PUMP POWERS ON NORMALLY WITH NO ALARMS, AND THERE WAS NO DAMAGE FOUND TO THE POWER CIRCUIT. NO DEFECTS WERE FOUND TO THE BATTERY CAP CONNECTIONS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OCCURRING.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT HE NOTICED THE PUMP SCREEN WAS BLANK. HE TOOK THE BATTERY OUT OF THE PUMP AND ATTEMPTED TO REBOOT; THE PUMP BEEPED BUT THE SCREEN REMAINED BLANK. HE TRIED TO INSERT A BATTERY FROM A DIFFERENT PACK AND AGAIN THE PUMP BEEPED BUT THE SCREEN REMAINED BLANK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 7 YR