UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2232163
·
Received September 2, 2011
Report
- Report Number
- 2050012-2011-04756
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Date of Event
- August 2, 2011
- Report Date
- August 2, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
SERVICE WAS NOT DISPATCHED FOR THIS EVENT AS THE CUSTOMER RESOLVED THE ISSUE. THE CUSTOMER CONTINUES TO MONITOR THE INSTRUMENT FOR ANY FURTHER ISSUES. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER WAS GENERATING "LOW PRESSURE - LOW" ERRORS AFTER CHANGING WASH CONCENTRATE. THE CUSTOMER STOPPED THE INSTRUMENT AND RESEATED THE WASH CONCENTRATE CAP; WASH CONCENTRATE WAS SPILLED ON THE FLOOR. THE CUSTOMER STOPPED THE INSTRUMENT AND ADJUSTED THE 10 PSI REGULATOR TO 10.3. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED IN THIS EVENT.