UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2232163 · Received September 2, 2011

Report

Report Number
2050012-2011-04756
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 2, 2011
Report Date
August 2, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS NOT DISPATCHED FOR THIS EVENT AS THE CUSTOMER RESOLVED THE ISSUE. THE CUSTOMER CONTINUES TO MONITOR THE INSTRUMENT FOR ANY FURTHER ISSUES. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER WAS GENERATING "LOW PRESSURE - LOW" ERRORS AFTER CHANGING WASH CONCENTRATE. THE CUSTOMER STOPPED THE INSTRUMENT AND RESEATED THE WASH CONCENTRATE CAP; WASH CONCENTRATE WAS SPILLED ON THE FLOOR. THE CUSTOMER STOPPED THE INSTRUMENT AND ADJUSTED THE 10 PSI REGULATOR TO 10.3. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600 PRO N/A