FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2232162 · Received September 2, 2011

Report

Report Number
2050012-2011-04760
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 1, 2011
Report Date
August 2, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER STATED THAT A WASH COLLAR VACUUM/WASTE VALVE WAS REMOVED OFF OF ANOTHER INSTRUMENT AND INSTALLED IT ON THE LEAKING INSTRUMENT WHICH RESOLVED THE ISSUE. SERVICE WAS NOT DISPATCHED FOR THIS EVENT. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE MODULAR CHEMISTRY (MC) SAMPLE PROBE WAS LEAKING FLUID IN THE UNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.