FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 800 SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2232162
·
Received September 2, 2011
Report
- Report Number
- 2050012-2011-04760
- Event Type
- Malfunction
- Date Received
- September 2, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 2, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER STATED THAT A WASH COLLAR VACUUM/WASTE VALVE WAS REMOVED OFF OF ANOTHER INSTRUMENT AND INSTALLED IT ON THE LEAKING INSTRUMENT WHICH RESOLVED THE ISSUE. SERVICE WAS NOT DISPATCHED FOR THIS EVENT. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE MODULAR CHEMISTRY (MC) SAMPLE PROBE WAS LEAKING FLUID IN THE UNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.