FDA Adverse Event Malfunction Summary report: N

INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 2232145 · Received September 1, 2011

Report

Report Number
9611451-2011-00540
Event Type
Malfunction
Date Received
September 1, 2011
Report Date
August 3, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K034026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WIL PROVIDE A FOLLOW UP REPORT UPON RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT236 INFANT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR INVESTIGATION. THE RETURNED DEVICE HAD BEEN OPENED BY THE CUSTOMER, REMOVED FROM ITS ORIGINAL PACKAGE AND PACKED IN A PLASTIC BAG. A VISUAL INSPECTION WAS PERFORMED. RESULTS: NO PHYSICAL DAMAGE WAS OBSERVED TO THE PRODUCT, BUT THE SWIVEL WAS MISSING FROM THE PRODUCT PACKAGING. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 110509. CONCLUSION: IT IS HIGHLY UNLIKELY THAT THE BREATHING CIRCUIT KIT WOULD HAVE LEFT THE PRODUCTION LINE WITHOUT A SWIVEL FITTED. ALL BREATHING CIRCUIT KITS ARE 100% PRESSURE TESTED AND A MISSING SWIVEL IN THE BREATHING CIRCUIT KIT WOULD HAVE CAUSED IT TO FAIL THE PRESSURE TEST. IT IS MOST LIKELY THAT THE SWIVEL WAS LOST POST PRODUCTION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT AN RT236 BREATHING CIRCUIT WAS MISSING THE Y-PIECE. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(4) REPORTED THAT AN RT236 BREATHING CIRCUIT WAS MISSING THE Y-PIECE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT236 110509

Patients

Seq Age Sex Outcome Treatment
1