FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2232130 · Received September 2, 2011

Report

Report Number
2050012-2011-04758
Event Type
Malfunction
Date Received
September 2, 2011
Date of Event
August 2, 2011
Report Date
August 2, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(6) 2011 AND REPLACED THE CO2 MEMBRANE, QUAD RINGS, AND ELECTRODE. ISE WAS CALIBRATED WITH SATISFACTORY RESULTS. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES PRIOR TO RETURNING IT INTO SERVICE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING THAT A LEAK WAS OBSERVED FROM THE FLOW CELL AFTER THE CARBON DIOXIDE (CO2) MEASURING ELECTRODE WAS REPLACED. CUSTOMER TECHNICAL SUPPORT (CTS) DETERMINED THAT THE CUSTOMER HAD NOT PUT A MEMBRANE ON THE CO2 ELECTRODE PRIOR TO INSTALLING THE ELECTRODE; CUSTOMER WAS NOT AWARE THAT A MEMBRANE WAS REQUIRED ON THE CO2 ELECTRODE AND ONLY INSTALLED A RETAINER AND QUADRING ON THE ELECTRODE. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600 PRO NA

Patients

Seq Age Sex Outcome Treatment
1