UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-04800
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 4, 2011
- Report Date
- August 4, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE CUSTOMER SHUT DOWN THE DXC UNIT AND REBOOTED IT. THE CUSTOMER ALSO TRACED THE LINE DOWN TO THE BLOCK AND CLEANED THE BLOCK. THE LEAKING CONTINUED. THE CUSTOMER THEN CLEANED THE PROBE, CHECKED THE WASH COLLAR AND FLUSHED THE BLOCK. THE MODULAR CHEMISTRY (MC) PROBE AND WASH COLLAR WERE STILL LEAKING. A SERVICE REQUEST WAS GENERATED.A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE (B)(6) 2011 AND FOUND MC WASTE VALVE AND ANOTHER VALVE BETWEEN WASTE SUMP AND WASTE EXIT SUMP WERE NOT WORKING AND REPLACED THEM. FSE RETURNED ON-SITE (B)(6) 2011 AFTER CUSTOMER NOTED THAT HYDRO WAS LEAKING AND FOUND THAT ONE VALVE WAS NOT PROPERLY SWITCHING BETWEEN PRESSURE AND VACUUM CAUSING THE MAIN WASTE SUMP TO BACK UP. FSE ALSO FOUND NO FOAM VALVE HAD FAILED, CONTRIBUTING TO THE PROBLEM. FSE REPLACED BOTH THE VALVES AND VERIFIED PROPER OPERATION OF THE WASTE SYSTEM AND CALIBRATED CHEMISTRIES AS NEEDED AND RAN QC. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE MODULAR CHEMISTRY (MC) SAMPLE PROBE AND WASH COLLAR IN THE UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM WERE NOT ASPIRATING PROPERLY AND WATER WAS RUNNING DOWN THE PROBE. NO INJURY OR EXPOSURE WAS REPORTED AND NO ERRONEOUS RESULTS WERE GENERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |